Title of article
Characterization of BSA unfolding and aggregation using a single-capillary viscometer and dynamic surface tension detector
Author/Authors
Bramanti، نويسنده , , Emilia and Ferrari، نويسنده , , Carlo and Angeli، نويسنده , , Valeria and Onor، نويسنده , , Massimo and Synovec، نويسنده , , Robert E.، نويسنده ,
Issue Information
ماهنامه با شماره پیاپی سال 2011
Pages
9
From page
2553
To page
2561
Abstract
A dynamic surface tension detector (DSTD) has been equipped with an additional pressure sensor for simultaneous viscosity measurements, as a detector for flow injection analysis. The viscosity measurement is based on a single capillary viscometer (SCV) placed in parallel configuration with the DSTD. The viscometer in the optimized conditions consists of a PEEK capillary (i.d. = 0.25 mm, L = 75 cm) kept at constant temperature using a thermostatic bath, which leads on the two sides to the two arms of a differential piezoelectric pressure transducer with a range of 0–35 psi. The DSTD, described previously, measures the changing pressure across the liquid/air interface of 2 μL drops repeatedly forming at the end of a capillary.
rformance has been evaluated by measuring dynamic viscosity of water/glycerol mixtures analysed in flow injection and comparing the results with the values reported in the literature. The detection limits of SCV and DSTD, calculated as 3σ of the blank, were 0.012 cP and 0.6 dyn cm−1, respectively.
-SCV-DSTD system has been applied to the study of temperature-induced denaturation/aggregation process in bovine serum albumin (BSA).
sults have been supported and discussed with respect to BSA conformational analysis performed using Fourier Transform infrared spectroscopy.
Keywords
DSTD , ?-Aggregation , denaturation , FTIR conformational analysis , Bovine serum albumin , Proteins , VISCOSITY , Surface Tension
Journal title
Talanta
Serial Year
2011
Journal title
Talanta
Record number
1663488
Link To Document