Title of article
Sputtered bismuth screen-printed electrode: A promising alternative to other bismuth modifications in the voltammetric determination of Cd(II) and Pb(II) ions in groundwater
Author/Authors
Sosa، نويسنده , , Velia and Serrano، نويسنده , , Nْria and Ariٌo، نويسنده , , Cristina and Dيaz-Cruz، نويسنده , , José Manuel and Esteban، نويسنده , , Miquel، نويسنده ,
Issue Information
ماهنامه با شماره پیاپی سال 2014
Pages
5
From page
348
To page
352
Abstract
A commercially available sputtered bismuth screen-printed electrode (BispSPE) has been pioneeringly applied for the simultaneous determination of Cd(II) and Pb(II) ions in a certified groundwater sample by means of differential pulse anodic stripping voltammetry (DPASV) as an alternative to more conventional bismuth screen-printed carbon electrodes (BiSPCEs). BispSPEs can be used for a large set of measurements without any previous plating or activation. The obtained detection and quantification limits suggest that BispSPEs produce a better analytical performance as compared to In-situ BiSPCE for Pb(II) and Cd(II) determination, but also to Ex-situ BiSPCE for Cd(II) determination. The results confirm the applicability of these devices for the determination of low level concentrations of these metal ions in natural samples with very high reproducibility (0.7% and 2.5% for Pb(II) and Cd(II) respectively), and good trueness (0.3% and 2.4% for Pb(II) and Cd(II) respectively).
Keywords
Bismuth sputtered electrodes , Metal analysis , Stripping voltammetry , Bismuth screen-printed electrodes
Journal title
Talanta
Serial Year
2014
Journal title
Talanta
Record number
1669894
Link To Document