Title of article
A simple recirculating flow system for the calibration of polar organic chemical integrative samplers (POCIS): Effect of flow rate on different water pollutants
Author/Authors
Di Carro، نويسنده , , Marina and Bono، نويسنده , , Luca and Magi، نويسنده , , Emanuele، نويسنده ,
Issue Information
ماهنامه با شماره پیاپی سال 2014
Pages
4
From page
30
To page
33
Abstract
A calibration system for POCIS was developed and used to calculate the sampling rates of eight analytes belonging to pesticides, non-steroidal anti-inflammatory drugs and perfluorinated compounds: atrazine, propazine, terbutylazine, diclofenac, ibuprofen, ketoprofen, perfluorooctanoic acid and perfluorooctanesulfonate. Experiments with a linear velocity of 2.0, 5.1, 10.2 and 15.3 cm/s were carried out for 96 h using two different analyte concentrations. POCIS extracts were analyzed by liquid chromatography–tandem mass spectrometry (LC–MS/MS), using multiple reaction monitoring to maximize sensitivity. Results highlighted that the calculated sampling rates are rather constant at the considered concentrations and flow rates. Obtained values of sampling rates were then employed to calculate Time-Weighted Average concentration of the analytes in river and drinking waters.
Keywords
sampling rate , Time-Weighted Average concentration (TWA) , Polar organic chemical integrative sampler (POCIS) , Polar organic micropollutant , Water monitoring
Journal title
Talanta
Serial Year
2014
Journal title
Talanta
Record number
1670055
Link To Document