Title of article
Determination of traces of cobalt in the presence of nioxime and cetyltrimethylammonium bromide by adsorptive stripping voltammetry
Author/Authors
Korolczuk، نويسنده , , Mieczyslaw and Moroziewicz، نويسنده , , Aleksandra and Grabarczyk، نويسنده , , Malgorzata and Paluszek، نويسنده , , Katarzyna، نويسنده ,
Issue Information
ماهنامه با شماره پیاپی سال 2005
Pages
5
From page
1003
To page
1007
Abstract
A sensitive method of Co(II) determination by adsorptive stripping voltammetry is presented. The method exploits the enhancement of cobalt peak current observed in the system Co(II)-nioxime-cetyltrimethylammonium bromide-piperazine-N,N′-bis(2-ethanesulfonic acid). The calibration plot for an accumulation time of 60 s is linear from 5 × 10−11 to 3 × 10−9 mol L−1. The relative standard deviation is 3.8% for Co(II) determination at concentration 1 × 10−9 mol L−1. The detection limit is 1.7 × 10−11 mol L−1. The validation of the method is performed by the analyses of certified reference materials and comparing the result of Co(II) determination in river water sample by the proposed method with those obtained by ET AAS. The main advantage of this new system is the micro-trace Co(II) determination by adsorptive stripping voltammetry, as compared to those described before, a low concentration of the supporting electrolyte used, and so commercially available reagents without additional purification can be used.
Keywords
Cobalt , Determination , Adsorptive stripping voltammetry
Journal title
Talanta
Serial Year
2005
Journal title
Talanta
Record number
1674092
Link To Document