• Title of article

    Comparison between the performances of amperometric immunosensors for cholera antitoxin based on three enzyme markers

  • Author/Authors

    Ionescu، نويسنده , , Rodica E. and Gondran، نويسنده , , Chantal and Cosnier، نويسنده , , Serge and Gheber، نويسنده , , Levi A. and Marks، نويسنده , , Robert S.، نويسنده ,

  • Issue Information
    ماهنامه با شماره پیاپی سال 2005
  • Pages
    6
  • From page
    15
  • To page
    20
  • Abstract
    We developed a novel copolymer modified amperometric immunosensor for the detection of cholera antitoxin (anti-CT), by the electropolymerization of pyrrole–biotin and pyrrole–lactitobionamide monomers on platinum or glassy carbon electrodes. In the detection of cholera antitoxin we have used three enzymatic marker detection systems based on HRP-labeled rabbit IgG antibodies, biotinylated polyphenol-oxidase (PPO-B) and biotinylated glucose-oxidase (GOX-B). The comparison of the electro-enzymatic performances of these three configurations with different substrates, clearly shows that the more sensitive amperometric immunosensor was based on HRP with a lower limit of detection of 50 ng/ml anti-CT using hydroquinone/H2O2 system. The response time for this substrate was in range of 5–30 s. The HRP-amperometric immunosensor has thus proven to be a very sensitive tool to monitor nanomolar concentrations of anti-CT.
  • Keywords
    Poly(pyrrole–biotin , pyrrole–lactitobionamide) , Anti-cholera toxin antibody , Amperometric immunosensors , Permeability
  • Journal title
    Talanta
  • Serial Year
    2005
  • Journal title
    Talanta
  • Record number

    1674183