• Title of article

    Sampled voltammetry on an electrode array for the renewal of the electrode surface and the analytical solution during the analysis

  • Author/Authors

    Mignard، نويسنده , , Laurence and Denoual، نويسنده , , Matthieu and Lavastre، نويسنده , , Olivier and Floner، نويسنده , , Didier and Geneste، نويسنده , , Florence، نويسنده ,

  • Issue Information
    روزنامه با شماره پیاپی سال 2013
  • Pages
    5
  • From page
    83
  • To page
    87
  • Abstract
    Polarography with dropping mercury electrode has been widely used in electroanalysis. However, the method is less and less employed due to the toxicity of mercury. In this work, we have shown that it is possible to replace the dropping electrode by a working electrode array, allowing the renewal of the electrode surface and of the analytical solution during the analysis. This new concept has been demonstrated on copper analysis. Sampled current voltammetry has been carried out on an electrode array, giving rise to I vs. E curves with a limiting diffusion plateau. The principle can be extended to other electroanalytical methods as exemplified here with differential pulse voltammetry. Linear calibration curves have been obtained with both methods and a limit of detection of 2 × 10−5 mol L−1 has been reached for copper detection by differential pulse voltammetry.
  • Keywords
    Electrode renewal , Electroanalysis , array , DPV , Copper
  • Journal title
    Journal of Electroanalytical Chemistry
  • Serial Year
    2013
  • Journal title
    Journal of Electroanalytical Chemistry
  • Record number