• Title of article

    Sensitivity improvement for antimony determination by using in-situ atom trapping in a slotted quartz tube and flame atomic absorption spectrometry

  • Author/Authors

    Sezgin and Titretir، نويسنده , , Serap and ??k، نويسنده , , Ahmet ?nanç and Arslan، نويسنده , , Yasin and Ataman، نويسنده , , O. Yavuz، نويسنده ,

  • Issue Information
    روزنامه با شماره پیاپی سال 2012
  • Pages
    6
  • From page
    63
  • To page
    68
  • Abstract
    Significant improvement has been achieved for antimony determination using a slotted quartz tube (SQT) as an atom trap (AT) for in situ preconcentration and flame atomic absorption spectrometry (FAAS). The suggested technique consists of trapping analyte species during ordinary nebulization followed by releasing the collected analyte via introducing organic solvent. Procedures and analytical figures of merit have been presented for the techniques called FAAS, SQT-FAAS and finally SQT-AT-FAAS with the relevant comparisons. Analytical parameters, namely composition of the aqueous medium, sample flow rate, flame conditions, distance between burner head and SQT, sampling period and type of organic solvent and its volume have been optimized. Using SQT-AT-FAAS, a sensitivity enhancement of 369 fold has been obtained, 3 s limit of detection was 3.9 μg L− 1 when 25.0 mL of sample was collected in 4.0 min. Interference effects of some elements on antimony signal were studied.
  • Keywords
    Antimony , Flame atomic absorption spectrometry , Slotted quartz tube , Atom trap
  • Journal title
    Spectrochimica Acta Part B Atomic Spectroscopy
  • Serial Year
    2012
  • Journal title
    Spectrochimica Acta Part B Atomic Spectroscopy
  • Record number