Title of article
Comparison of laser induced breakdown spectroscopy and spark induced breakdown spectroscopy for determination of mercury in soils
Author/Authors
Srungaram، نويسنده , , Pavan K. and Ayyalasomayajula، نويسنده , , Krishna K. and Yu-Yueh، نويسنده , , Jin-Fang and Singh، نويسنده , , Jagdish P.، نويسنده ,
Issue Information
روزنامه با شماره پیاپی سال 2013
Pages
6
From page
108
To page
113
Abstract
Mercury is a toxic element found throughout the environment. Elevated concentrations of mercury in soils are quite hazardous to plants growing in these soils and also the runoff of soils to nearby water bodies contaminates the water, endangering the flora and fauna of that region. This makes continuous monitoring of mercury very essential. This work compares two potential spectroscopic methods (laser induced breakdown spectroscopy (LIBS) and spark induced breakdown spectroscopy (SIBS)) at their optimum experimental conditions for mercury monitoring. For LIBS, pellets were prepared from soil samples of known concentration for generating a calibration curve while for SIBS, soil samples of known concentration were used in the powder form. The limits of detection (LODs) of Hg in soil were calculated from the Hg calibration curves. The LOD for mercury in soil calculated using LIBS and SIBS is 483 ppm and 20 ppm, respectively. The detection range for LIBS and SIBS is discussed.
Keywords
atomic emission spectroscopy , Spark Induced Breakdown Spectroscopy , Laser induced breakdown spectroscopy
Journal title
Spectrochimica Acta Part B Atomic Spectroscopy
Serial Year
2013
Journal title
Spectrochimica Acta Part B Atomic Spectroscopy
Record number
1688862
Link To Document