Title of article
Sensitive high-performance liquid chromatographic assay for motexafin gadolinium and motexafin lutetium in human plasma
Author/Authors
Parise، نويسنده , , Robert A and Miles، نويسنده , , Dale R and Egorin، نويسنده , , Merrill J، نويسنده ,
Issue Information
روزنامه با شماره پیاپی سال 2000
Pages
8
From page
145
To page
152
Abstract
We present new HPLC methods for the quantitation in human plasma of two investigative metallotexaphyrin agents, motexafin gadolinium (Gd-Tex) and motexafin lutetium (Lu-Tex). Each assay uses: the other texaphyrin analogue as an internal standard; protein precipitation with acetonitrile:methanol (50:50, v/v); an ODS reversed-phase column; an isocratic mobile phase of 100 mM ammonium acetate, pH 4.3:acetonitrile:methanol (59:21:20, v/v/v); and absorbance detection at 470 nm. The Gd-Tex assay has a lower limit of quantitation (LLOQ) of 0.01 μM and is linear between 0.01and 30 μM. The Lu-Tex assay has an LLOQ of 0.1 μM and is linear between 0.1 and 30 μM. The assays are suited for in vivo preclinical studies and clinical trials because they require minimal amounts of plasma, are sensitive, and involve a 30-min run time. These assays are important tools for evaluating the potential of Gd-Tex and Lu-Tex as a radiation enhancer and photosensitizer, respectively.
Keywords
Motexafin gadolinium , Texaphyrin , Motexafin lutetium , Lanthanide
Journal title
Journal of Chromatography B Biomedical Sciences and Applications
Serial Year
2000
Journal title
Journal of Chromatography B Biomedical Sciences and Applications
Record number
1704212
Link To Document