• Title of article

    Reversed-phase high-performance liquid chromatography of nicotinic acid mononucleotide for measurement of quinolinate phosphoribosyltransferase

  • Author/Authors

    Shibata، نويسنده , , Katsumi and Fukuwatari، نويسنده , , Tsutomu and Sugimoto، نويسنده , , Etsuro، نويسنده ,

  • Issue Information
    روزنامه با شماره پیاپی سال 2000
  • Pages
    5
  • From page
    281
  • To page
    285
  • Abstract
    A system has been developed for the determination of quinolinate phosphoribosyltransferase (QPRT) activity in liver and kidney homogenates using HPLC. A product, nicotinic acid mononucleotide (NaMN), is separated by reversed-phase chromatography (a Tosoh ODS 80TS was used as an analytical column) using a mixture of 10 mM KH2PO4–K2HPO4 buffer (pH 7.0) containing 1.48 g/l tetra-n-butylammonium bromide–acetonitrile (9:1, v/v) as a mobile phase. The flow-rate was 1.0 ml/min, the detection wavelength was 265 nm. The column temperature was maintained at 40°C. Under these conditions, NaMN was eluted at about 8.1 min. Sample preparation was very straightforward. The reaction mixture of QPRT assay was stopped by immersing the tube into a boiling water bath, the resulting supernatant was filtered, and the filtrate was directly injected into a HPLC system. The total HPLC analysis time was approximately 20 min.
  • Keywords
    quinolinate phosphoribosyltransferase , nicotinic acid mononucleotide , enzymes
  • Journal title
    Journal of Chromatography B Biomedical Sciences and Applications
  • Serial Year
    2000
  • Journal title
    Journal of Chromatography B Biomedical Sciences and Applications
  • Record number

    1704254