• Title of article

    Determination of albumin and myoglobin in dialysate and ultrafiltrate samples by high-performance size-exclusion chromatography

  • Author/Authors

    Liang، نويسنده , , Hairui and Scott، نويسنده , , Meri K. and Murry، نويسنده , , Daryl J. and Sowinski، نويسنده , , Kevin M.، نويسنده ,

  • Issue Information
    روزنامه با شماره پیاپی سال 2001
  • Pages
    11
  • From page
    141
  • To page
    151
  • Abstract
    A high-performance size-exclusion chromatographic method was developed, validated and implemented for simultaneous and quantitative determination of albumin and myoglobin along with inulin, vancomycin and creatinine in dialysate and ultrafiltrate samples from in vitro hemodialysis experiments. The experimental parameters including mobile phase pH, ionic strength, detection wavelength, flow-rate, injection volume were first optimized for the determination of albumin, myoglobin, inulin, vancomycin and creatinine. The peak height ratio and detection limits of the proteins were then comparatively studied at 210, 254 and 280 nm by UV and diode array detection. The method was further validated by evaluating the linearity, precision and accuracy of the proteins. The assay was finally implemented to the simultaneous and quantitative determination of the proteins in dialysate and ultrafiltrate samples.
  • Keywords
    Albumin , myoglobin , Proteins
  • Journal title
    Journal of Chromatography B Biomedical Sciences and Applications
  • Serial Year
    2001
  • Journal title
    Journal of Chromatography B Biomedical Sciences and Applications
  • Record number

    1704823