• Title of article

    Assessment of the ultrasonic dental scaler insert

  • Author/Authors

    Trenter، نويسنده , , S.C and Walmsley، نويسنده , , A.D and Landini، نويسنده , , G and Shippen، نويسنده , , J.M، نويسنده ,

  • Issue Information
    روزنامه با شماره پیاپی سال 2002
  • Pages
    6
  • From page
    139
  • To page
    144
  • Abstract
    The ultrasonic dental scaler is an invaluable tool in the fight against dental calculus and periodontal disease. The purpose of this study was to assess the natural frequency of the straight internal ultrasonic scaler insert 30K™ FSI®-SLI™-10S (Dentsply International, York, USA) and to measure the displacement observed at the tip when operated under different generator settings. Two approaches were used: finite element analysis and scanning laser vibrometry. The Finite Element Analysis method was employed to calculate the expected modes of vibration of the insert. Scanning laser vibrometry was used to measure displacements of the tip during operation. It was found the insert had two modes of vibration around the operating frequency of 30kHz; these modes were first order bending in two orthogonal planes.
  • Keywords
    Periodontology , Finite element analysis , Scanning laser vibrometry , Ultrasonic scaler
  • Journal title
    Medical Engineering and Physics
  • Serial Year
    2002
  • Journal title
    Medical Engineering and Physics
  • Record number