• Title of article

    Accuracy estimations of testing of infusion devices using weighing balances

  • Author/Authors

    Clarkson، نويسنده , , D.M، نويسنده ,

  • Issue Information
    روزنامه با شماره پیاپی سال 2002
  • Pages
    7
  • From page
    229
  • To page
    235
  • Abstract
    The role of gravimetric balances in the evaluation of infusion devices is considered from a perspective of accuracy for such evaluations and the actual physical and interface characteristics of such balances as specialist measurement devices. In particular, the dynamic properties of weighing balances are discussed in relation to observations of variation of response times and stability settings. This work has identified characteristics, which may influence flow measurement techniques as defined by measurement protocols adopted by relevant standards and those developed independently. While a dominant error is introduced due to balance resolution error, consideration of timing errors associated with data capture is also essential.
  • Keywords
    Flow measurement , Infusion devices , Weighing balance , Gravimetric techniques
  • Journal title
    Medical Engineering and Physics
  • Serial Year
    2002
  • Journal title
    Medical Engineering and Physics
  • Record number

    1727684