• Title of article

    Thermoluminescent dosimetry of the SymmetraTM 125I model I25.S06 interstitial brachytherapy seed

  • Author/Authors

    Williamson، Jeffrey F. نويسنده , , Patel، Neil S. نويسنده , , Chiu-Tsao، Sou-Tung نويسنده , , Fan، Pei نويسنده , , Duckworth، Tamara نويسنده , , Shasha، Daniel نويسنده , , Harrison، Louis B. نويسنده ,

  • Issue Information
    روزنامه با شماره پیاپی سال 2001
  • Pages
    -1760
  • From page
    1761
  • To page
    0
  • Abstract
    As the efficacy of brachytherapy prostate treatment is becoming realized, new models of 125I seeds are being introduced. In this article we present thermoluminescent dosimetry (TLD) in a solid water phantom for a new design of 125I seed (UroMed/Bebig SymmetraTM, Model I25.S06). TLD cubes, LiF TLD-100, from Bicron (Solon, OH) with dimension 1×1×1 mm3 were irradiated at various distances from the seed at angles ranging from 0° to 90° in 10° increments. The TLD detectors were calibrated by irradiation in a 60Co teletherapy beam. Monte Carlo simulation was used to account for TLD energy dependence and the deviation of solid water composition (as determined by chemical analysis of a sample) from liquid water. Dose rates per unit air kerma strength were determined based on calibrations traceable to the 1999 NIST standard (corrected for NIST measurement errors made in 1999) for the SymmetraTM seed. Dose data is presented in TG-43 format as a function of distance and angle. Values for lambda, F (r,theta), g(r), and the anisotropy constant are obtained for use in radiation treatment planning (RTP) software. The dose rate constant was determined to be 1.033±6.4% cGy h-1 U-1, which is comparable to model 6702 and higher than model 6711. We find the relative dose distributions of the SymmetraTM seed are similar to model 6702, and less anisotropic than model 6711. After accounting for deviation of measured solid water composition from the manufacturerʹs specification, good agreement between TLD results and Monte-Carlo-aided values was found.
  • Keywords
    power quality , Fault current limiter , short circuit current , transient over voltage
  • Journal title
    MEDICAL PHYSICS
  • Serial Year
    2001
  • Journal title
    MEDICAL PHYSICS
  • Record number

    1728