• Title of article

    An improved approach for measurement of stroke volume during treadmill exercise

  • Author/Authors

    Song، نويسنده , , C.G. and Kim، نويسنده , , D.W.، نويسنده ,

  • Issue Information
    روزنامه با شماره پیاپی سال 2003
  • Pages
    6
  • From page
    517
  • To page
    522
  • Abstract
    In this study, a new ensemble averaging technique was developed to measure cardiac output during treadmill exercise. Each dz/dt peak is used as a starting point for ensemble averaging, instead of the conventionally used R point of ECG, in order to prevent the peak dz/dt waveform (C point) to be smoothed, which results in increased dz/dtmax. When using the R point as a reference, time interval from the R point to the peak dz/dt varies for each heart beat. Signal to noise ratios (SNR) of the five subjects with the new technique were 12.9, 16. 4, 10.6, 14.0, and 4.6% higher than those with the conventional one at rest and during four levels of exercise, respectively. Power density spectra of the averaged dz/dt waveforms by the peak dz/dt and R point referencing were also analyzed. Stroke volumes of the five subjects averaged by the peak dz/dt were 4.9, 6.8, 7.6, 7.4, and 9.8% higher than those by the R point at rest and during four levels of exercise, respectively.
  • Keywords
    Stroke volume , Impedance cardiography , ensemble average
  • Journal title
    Medical Engineering and Physics
  • Serial Year
    2003
  • Journal title
    Medical Engineering and Physics
  • Record number