• Title of article

    A novel liner locking mechanism enhances retention stability

  • Author/Authors

    Macdonald، نويسنده , , Warren and Aspenberg، نويسنده , , Anders and Jacobsson، نويسنده , , Magnus and Carlsson، نويسنده , , Lars V.، نويسنده ,

  • Issue Information
    روزنامه با شماره پیاپی سال 2003
  • Pages
    8
  • From page
    747
  • To page
    754
  • Abstract
    Acetabular liner retention of a novel design of liner locking was evaluated in static and cyclic endurance modes. The locking mechanism combines geometric form and accurate machining to give high conformity to the acetabular shell and minimise relative motion against the metal shell, minimising debris generation and escape or ingress. Using amended test liners with integral coupling, mean static pullout strength was determined to be 399±53 N and lever-out strength 28.03±2.8 N m. Cyclic loading of 5 N m for up to 10 million cycles caused no significant reduction in strength, no detectable fretting wear, and the sealing mechanism prevented particle access between the cup interior and the “effective joint space”. The stability measured ensures secure and reliable in vivo retention of the liner, comparable with extant component designs using other liner locking mechanisms.
  • Keywords
    Acetabular implant , Polyethylene , mechanical testing , disassembly , Modular component
  • Journal title
    Medical Engineering and Physics
  • Serial Year
    2003
  • Journal title
    Medical Engineering and Physics
  • Record number

    1728098