• Title of article

    Finite element analysis of a femoral retrograde intramedullary nail subject to gait loading

  • Author/Authors

    Cheung، نويسنده , , G and Zalzal، نويسنده , , P and Bhandari، نويسنده , , M and Spelt، نويسنده , , J.K and Papini، نويسنده , , M، نويسنده ,

  • Issue Information
    روزنامه با شماره پیاپی سال 2004
  • Pages
    16
  • From page
    93
  • To page
    108
  • Abstract
    Intramedullary nails are routinely used in the treatment of fractures of the femur. While their effectiveness has been demonstrated clinically, a number of complications, including bone refracture and implant failure, persist. This paper presents novel three-dimensional finite element (FE) models, at four stages of gait, of: (i) a realistic femur analogue known as third generation composite bone, and (ii) a system consisting of an intramedullary nail implanted in the femur of (i). A comparison of experimentally measured strains on the surface of the femur with those predicted by the FE model revealed good agreement. The models were then used to identify implant/bone load sharing patterns, and areas of stress concentration in both the intramedullary nail and the bone, when statically locked by one or two screws at either end. The results of this study can be used to guide future implant design and surgical procedure.
  • Keywords
    Fracture fixation , Biomechanics , Finite element analysis , Intramedullary nail , Femur , STRESS
  • Journal title
    Medical Engineering and Physics
  • Serial Year
    2004
  • Journal title
    Medical Engineering and Physics
  • Record number