Title of article
Validation of a numerical 3-D fluid–structure interaction model for a prosthetic valve based on experimental PIV measurements
Author/Authors
Guivier-Curien، نويسنده , , Carine and Deplano، نويسنده , , Valérie and Bertrand، نويسنده , , Eric، نويسنده ,
Issue Information
روزنامه با شماره پیاپی سال 2009
Pages
8
From page
986
To page
993
Abstract
A numerical 3-D fluid–structure interaction (FSI) model of a prosthetic aortic valve was developed, based on a commercial computational fluid dynamics (CFD) software program using an Arbitrary Eulerian Lagrangian (ALE) formulation. To make sure of the validity of this numerical model, an equivalent experimental model accounting for both the geometrical features and the hydrodynamic conditions was also developed. The leaflet and the flow behaviours around the bileaflet valve were investigated numerically and experimentally by performing particle image velocimetry (PIV) measurements. Through quantitative and qualitative comparisons, it was shown that the leaflet behaviour and the velocity fields were similar in both models. The present study allows the validation of a fully coupled 3-D FSI numerical model. The promising numerical tool could be therefore used to investigate clinical issues involving the aortic valve.
Keywords
Bileaflet prosthetic valve , Fluid–structure interaction , Particle image velocimetry
Journal title
Medical Engineering and Physics
Serial Year
2009
Journal title
Medical Engineering and Physics
Record number
1730697