• Title of article

    Modelization of a self-opening peripheral neural interface: A feasibility study

  • Author/Authors

    Cutrone، نويسنده , , Annarita and Sergi، نويسنده , , Pier Nicola and Bossi، نويسنده , , Silvia and Micera، نويسنده , , Silvestro، نويسنده ,

  • Issue Information
    روزنامه با شماره پیاپی سال 2011
  • Pages
    8
  • From page
    1254
  • To page
    1261
  • Abstract
    In this paper a self-opening intrafascicular neural interface (SELINE) has been modeled using both a theoretical approach and a Finite Element (FE) analysis. This innovative self opening interface has several potential advantages such as: higher selectivity due to its three-dimensional structure and efficient anchorage system. Mechanical, structural and micro-technological issues have been considered to obtain an effective design of the electrode, as a feasibility study of the self-opening approach. A simple framework has been provided to model the insertion and partial retraction into peripheral nerves, resulting in the opening of wings. ntegrated approach results in a rational procedure to optimize kinematics, geometry, and structural properties of peripheral interfaces. The design and feasibility study carried out in this work can potentially assure a correct behavior and dimensioning of the neural interface: in this way anomalous breakage should be avoided while mechanical and geometrical biocompatibility should increase.
  • Keywords
    Neural interface , Structural and FE analysis , peripheral nerve , Insertion procedure , Asymmetric kinematics
  • Journal title
    Medical Engineering and Physics
  • Serial Year
    2011
  • Journal title
    Medical Engineering and Physics
  • Record number