• Title of article

    Experimental investigation of the reciprocating ball pump (RBP)

  • Author/Authors

    Lieu، نويسنده , , C.F. and Chan، نويسنده , , W.K. and Ooi، نويسنده , , K.T.، نويسنده ,

  • Issue Information
    روزنامه با شماره پیاپی سال 2012
  • Pages
    8
  • From page
    1101
  • To page
    1108
  • Abstract
    This paper presents experimental investigations of a novel miniature reciprocating ball pump (RBP) as a potential infusion pump or drug delivery system. The objectives of this study are to validate the RBPʹs operational principle and to evaluate its performance. In the tests, the basic functionality of the pump has been established at different pumping frequencies ranging from 1.667 Hz (100 rpm) to 4.167 Hz (250 rpm) against different pressure heads ranging from 0 mmH2O and 100 mmH2O. In addition, experimental results show that the RBP can deliver a higher flow rate with a volumetric efficiency of 120% as compared to a conventional single acting reciprocating pump running under the same operating conditions.
  • Keywords
    Infusion pump , Reciprocating pump , Drug delivery system , Induced flow , Reciprocating ball pump