Title of article
Experimental investigation of the reciprocating ball pump (RBP)
Author/Authors
Lieu، نويسنده , , C.F. and Chan، نويسنده , , W.K. and Ooi، نويسنده , , K.T.، نويسنده ,
Issue Information
روزنامه با شماره پیاپی سال 2012
Pages
8
From page
1101
To page
1108
Abstract
This paper presents experimental investigations of a novel miniature reciprocating ball pump (RBP) as a potential infusion pump or drug delivery system. The objectives of this study are to validate the RBPʹs operational principle and to evaluate its performance. In the tests, the basic functionality of the pump has been established at different pumping frequencies ranging from 1.667 Hz (100 rpm) to 4.167 Hz (250 rpm) against different pressure heads ranging from 0 mmH2O and 100 mmH2O. In addition, experimental results show that the RBP can deliver a higher flow rate with a volumetric efficiency of 120% as compared to a conventional single acting reciprocating pump running under the same operating conditions.
Keywords
Infusion pump , Reciprocating pump , Drug delivery system , Induced flow , Reciprocating ball pump
Journal title
Medical Engineering and Physics
Serial Year
2012
Journal title
Medical Engineering and Physics
Record number
1731790