Abstract
Load deflection and hysteresis measurements were made on 37 wheelchair seating cushions according to ISO 16840-2:2007. Load deflection plots for all 37 cushions are reported and fundamental aspects of graph interpretation discussed. ISO hysteresis data are also reported and interpretation discussed.
Keywords
Load deflection , Mechanical Characteristics , Pressure , pressure ulcer , seating , Standard , Tissue integrity , Cushion , hysteresis , ISO , wheelchair , ISO 16840 , ISO 16840-2 , Foam