Magic-echo phase-encoding solid imaging with improved time resolution
Author/Authors
Hafner، نويسنده , , S. R. Barth، نويسنده , , P.، نويسنده ,
Issue Information
روزنامه با شماره پیاپی سال 1995
Pages
5
From page
441
To page
445
Abstract
The magic-echo phase-encoding solid imaging (MEPSI) sequence is based on the magic-echo pulse sequence for the refocusing of the dipolar interaction and on pure phase-encoding for spatial resolution in all dimensions. MEPSI has many advantages; however, as a consequence of pure phase-encoding, it is very time-consuming. In the following, two strategies are proposed to decrease the time necessary for the acquisition of an image by a factor of 2 per dimension.
Keywords
Magic-echo phase-encoding solid imaging (MEPSI) , Phase-encoding , Three-Dimensional