• Title of article

    A design methodology for short, whole-body, shielded gradient coils for MRI

  • Author/Authors

    Crozier، نويسنده , , Stuart and Doddrell، نويسنده , , David M.، نويسنده ,

  • Issue Information
    روزنامه با شماره پیاپی سال 1995
  • Pages
    6
  • From page
    615
  • To page
    620
  • Abstract
    A series of designs is presented for restricted length, whole-body, shielded gradient coils. By using the real space optimization technique, simulated annealing (SA), it is possible to produce viable gradient sets with a length-to-diameter ratio (LDR) of just 1.0. Radially remote return paths for the transverse coils aid in producing such short coils. While the linear regions of such coils cannot be as large as longer coils, they produce homogeneous linear regions suitable for use in whole-body imaging. The coil sets are well shielded even at such small LDRs.
  • Keywords
    Whole-body imaging , Whole-body , shielded gradient coils , Gradient coils , Simulated annealing (SA)
  • Journal title
    Magnetic Resonance Imaging
  • Serial Year
    1995
  • Journal title
    Magnetic Resonance Imaging
  • Record number