• Title of article

    An interleaved sequence for accurate and reproducible clinical measurement of Magnetization Transfer Ratio

  • Author/Authors

    Barker، نويسنده , , G.J. and Tofts، نويسنده , , P.S. and Gass، نويسنده , , A.، نويسنده ,

  • Issue Information
    روزنامه با شماره پیاپی سال 1996
  • Pages
    9
  • From page
    403
  • To page
    411
  • Abstract
    We demonstrate an interleaved dual spin echo-based sequence for quantitative measurement of Magnetisation Transfer Ratio (MTR) in a clinical environment that overcomes the problems of patient motion between scans faced by noninterleaved methods. The sequence also provides proton density and T2-weighted images, allowing direct comparison among the three contrast regimes. Phantom studies and in vivo measurements on normal controls show the sequence to be robust in normal use. The values of MTR calculated from the sequence are shown to be precise and reproducible enough to allow regional variations to be identified within and between white matter and other brain tissues.
  • Keywords
    Interleaving , Spin echo , magnetization transfer
  • Journal title
    Magnetic Resonance Imaging
  • Serial Year
    1996
  • Journal title
    Magnetic Resonance Imaging
  • Record number