• Title of article

    Image registration of MR and CT images using a frameless fiducial marker system

  • Author/Authors

    Kremser، نويسنده , , Christian and Plangger، نويسنده , , Clemens and Bِsecke، نويسنده , , Robert and Pallua، نويسنده , , Anton and Aichner، نويسنده , , Franz and Felber، نويسنده , , Stephan R.، نويسنده ,

  • Issue Information
    روزنامه با شماره پیاپی سال 1997
  • Pages
    7
  • From page
    579
  • To page
    585
  • Abstract
    A new system of fiducial stereotactic markers that can easily be adapted to various imaging modalities without losing image registration was developed and tested. Utilizing MR and CT imaging the accuracy of the new system was evaluated with phantom studies and preliminary patient studies. The markers are clearly visible without artifacts on both imaging modalities. The clear delineation of the marker dots on the images enables an accurate automated marker detection. Using the marker system, image registration was found to yield an accuracy of up to 1 mm, depending on the imaging modality and the employed marker arrangement. The presented marker system shall improve patient comfort in comparison to conventional fixed stereotactic frames if repeated, highly accurate registrations are necessary over longer periods.
  • Keywords
    Frameless stereotaxy , magnetic resonance , image registration , computed tomography
  • Journal title
    Magnetic Resonance Imaging
  • Serial Year
    1997
  • Journal title
    Magnetic Resonance Imaging
  • Record number

    1829353