• Title of article

    A four-element phased array coil for high resolution and parallel MR imaging of the knee

  • Author/Authors

    Kwok، نويسنده , , Wingchi E and Zhong، نويسنده , , Jianhui and You، نويسنده , , Zhigang and Seo، نويسنده , , Gwysuk and Totterman، نويسنده , , Saara M.S، نويسنده ,

  • Issue Information
    روزنامه با شماره پیاپی سال 2003
  • Pages
    7
  • From page
    961
  • To page
    967
  • Abstract
    A four-element phased array coil for MR imaging of the knee was designed, built and tested for clinical use at 1.5 Tesla. In routine imaging, it provides over twofold increase in signal-to-noise (SNR) compared to two commercially available knee coils, and supports higher spatial image resolution. The phased array knee coil was also tested for its compatibility with parallel MR imaging that reduces imaging time by several folds over conventional MR technique. Results obtained using SiMultaneous Acquisition of Spatial Harmonics (SMASH) technique shows that our phased array knee coil can be used with parallel MR imaging. These improvements may enhance knee diagnosis with higher image quality and reduced scan time.
  • Keywords
    Magnetic Resonance Imaging , Knee imaging , High-resolution imaging , Phased array coil , parallel imaging
  • Journal title
    Magnetic Resonance Imaging
  • Serial Year
    2003
  • Journal title
    Magnetic Resonance Imaging
  • Record number

    1831705