• Title of article

    Experimental development of a petal resonator surface coil

  • Author/Authors

    Rodrيguez، نويسنده , , Alfredo Odon and Hidalgo، نويسنده , , Sandra Silvia and Rojas، نويسنده , , Rafael and Barrios، نويسنده , , Fernando Alejandro، نويسنده ,

  • Issue Information
    روزنامه با شماره پیاپی سال 2005
  • Pages
    7
  • From page
    1027
  • To page
    1033
  • Abstract
    A surface coil for MRI was designed and built based on the principles of the petal resonator proposed by Mansfield [J Phys D Appl Phys 21 (1988) 1643]. This resonator coil design was named the petal resonator surface (PERES) coil and is composed of an eight-petal coil array and a central circular coil. A minimum separation of three times the petal coil radius is necessary to significantly decrease the mutual inductance. An analytical function for the PERES Signal-to-noise ratio (SNR) is obtained based on the quasistatic method. Theoretical plots of SNR enhancement yielded 26% and 35% more SNR over the circular coil and phased-array coils. Imaging experiments were first performed using a spectroscopy phantom on a 1.5-T commercial imager. Subsequently, brain images of healthy volunteers were obtained. Clinical MR imager compatibility allows this resonator coil to be used with conventional pulse sequences and imaging protocols. This coil design offers a new alternative to existing surface coils because it significantly increases the SNR.
  • Keywords
    SNR improvement , Brain , Surface resonator coil
  • Journal title
    Magnetic Resonance Imaging
  • Serial Year
    2005
  • Journal title
    Magnetic Resonance Imaging
  • Record number