• Title of article

    Sensitivity enhancement and compensation of RF penetration artifact with planar actively detunable quadrature surface coil

  • Author/Authors

    Peshkovsky، نويسنده , , Alexey and Kennan، نويسنده , , Richard P. and Nagel، نويسنده , , Ronald L. and Avdievich، نويسنده , , Nikolai I.، نويسنده ,

  • Issue Information
    روزنامه با شماره پیاپی سال 2006
  • Pages
    7
  • From page
    81
  • To page
    87
  • Abstract
    An actively detunable planar quadrature surface coil for human body imaging at 4 T has been constructed and compared with a conventional linear surface coil. The coil could be used as a transmit/receive or a receive-only device in combination with a volume transmit coil. Transmission, reception profiles and the corresponding images acquired with each coil, as well as with both individual modes of the quadrature coil, are presented. Data collected using a tissue equivalent loaded phantom recorded with the linear surface coil demonstrated significant intensity distortions due to RF penetration artifact. The quadrature surface coil, on the other hand, provided compensation of the artifact, separately in its transmission and reception profiles as well as in the resultant images. Substantial sensitivity gain was also observed for the quadrature coil compared to the linear device. Significant advantages of using the quadrature surface coil over the linear device at 4 T have, therefore, been demonstrated.
  • Keywords
    MRI , Quadrature surface coil , RF penetration artifact , High-field body imaging
  • Journal title
    Magnetic Resonance Imaging
  • Serial Year
    2006
  • Journal title
    Magnetic Resonance Imaging
  • Record number