Title of article
Cyclic variation of T1 in the myocardium at 3 T
Author/Authors
Wansapura، نويسنده , , Janaka and Gottliebson، نويسنده , , William and Crotty، نويسنده , , Eric and Fleck، نويسنده , , Robert، نويسنده ,
Issue Information
روزنامه با شماره پیاپی سال 2006
Pages
5
From page
889
To page
893
Abstract
Standard methods of longitudinal relaxation (T1) measurements in the heart produce only one T1 map of the myocardium, usually at the end diastole (ED). In this article, we investigated the feasibility of using a dual flip angle fast gradient echo technique in the steady state to generate a movie of T1 maps in the myocardium during the cardiac cycle. The effects of nonideal slice profile and transient steady state on the T1 measurements were evaluated by Bloch simulations. Based on these results, we introduce a linear correction to the measured T1 values, which was validated by phantom experiments. In vivo T1 cine maps in healthy volunteers show 70±7% drop in T1 from the ED to the end systole in the septum and a 43±13% drop in the left ventricular lateral wall. With further improvements, this technique could be used to assess the myocardial blood volume changes during the cardiac cycle.
Keywords
Cardiac MRI , t1 , myocardium
Journal title
Magnetic Resonance Imaging
Serial Year
2006
Journal title
Magnetic Resonance Imaging
Record number
1832318