• Title of article

    Fast multivoxel two-dimensional spectroscopic imaging at 3 T

  • Author/Authors

    Kim، نويسنده , , Dong-Hyun and Henry، نويسنده , , Roland and Spielman، نويسنده , , Daniel M.، نويسنده ,

  • Issue Information
    روزنامه با شماره پیاپی سال 2007
  • Pages
    7
  • From page
    1155
  • To page
    1161
  • Abstract
    The utility of multivoxel two-dimensional chemical shift imaging in the clinical environment will ultimately be determined by the imaging time and the metabolite peaks that can be detected. Different k-space sampling schemes can be characterized by their minimum required imaging time. The use of spiral-based readout gradients effectively reduces the minimum scan time required due to simultaneous data acquisition in three k-space dimensions (kx, ky and kf2). A 3-T spiral-based multivoxel two-dimensional spectroscopic imaging sequence using the PRESS excitation scheme was implemented. Good performance was demonstrated by acquiring preliminary in vivo data for applications, including brain glutamate imaging, metabolite T2 quantification and high-spatial-resolution prostate spectroscopic imaging. All protocols were designed to acquire data within a 17-min scan time at a field strength of 3 T.
  • Keywords
    Magnetic Resonance Spectroscopic Imaging , Spiral readout gradients , 2D J-resolved spectroscopy
  • Journal title
    Magnetic Resonance Imaging
  • Serial Year
    2007
  • Journal title
    Magnetic Resonance Imaging
  • Record number

    1832592