• Title of article

    A simplified formula for T1 contrast optimization for short-TR steady-state incoherent (spoiled) gradient echo sequences

  • Author/Authors

    Neelavalli، نويسنده , , Jaladhar and Haacke، نويسنده , , E. Mark، نويسنده ,

  • Issue Information
    روزنامه با شماره پیاپی سال 2007
  • Pages
    5
  • From page
    1397
  • To page
    1401
  • Abstract
    There are certain instances in practical magnetic resonance imaging where T1 changes by a small amount relative to a neighboring pixel or between scans for a given pixel. The source of these small changes in T1 can be caused either by changes in tissue water content or by the uptake of a contrast agent. For short repetition time (TR) spoiled gradient echo imaging, we show that a robust and a simple, easy to use back-of-the-envelope expression for the flip angle that optimizes contrast under these conditions is given by √3θE in radians or (180/π)√6TR/T1 in degrees. We show that for a TR/T1 ratio of up to 0.3 and for a T1 change of up to ±50%, this approximation to the optimal flip angle produces a contrast to within 6% of the theoretical maximum value and that these predictions are in good agreement with experiment.
  • Keywords
    FLASH , Ernst angle , RF spoiled gradient echo imaging , Optimal contrast
  • Journal title
    Magnetic Resonance Imaging
  • Serial Year
    2007
  • Journal title
    Magnetic Resonance Imaging
  • Record number

    1832619