Quantitative T1ρ imaging using phase cycling for B0 and B1 field inhomogeneity compensation
Author/Authors
Chen، نويسنده , , Weitian and Takahashi، نويسنده , , Atsushi and Han، نويسنده , , Eric، نويسنده ,
Issue Information
روزنامه با شماره پیاپی سال 2011
Pages
12
From page
608
To page
619
Abstract
T1ρ imaging is useful in a number of clinical applications. T1ρ preparation methods, however, are sensitive to non-uniformities of the B0 magnetic field and the B1 RF field. These common system imperfections can result in image artifacts and quantification errors in T1ρ imaging. We report on a phase-cycling method which can eliminate B1 RF inhomogeneity effects in T1ρ imaging. This method does not only correct for image artifacts but also for T2ρ contamination caused by B1 RF inhomogeneity. The presence of B0 magnetic field inhomogeneity can compromise the effectiveness of this method for B1 RF inhomogeneity correction. We demonstrate that, by combining the spin-locking scheme reported by Dixon et al. (Myocardial suppression in vivo by spin locking with composite pulses. Magn Reson Med 1996; 36:90–94) with phase cycling, we can simultaneously correct B0 magnetic field inhomogeneity effects and B1 RF inhomogeneity effects in T1ρ imaging. Phantom and in vivo data sets are used to demonstrate the proposed methods and to compare them with other existing T1ρ preparation methods.
Keywords
T1? , Off-resonance , B1 RF inhomogeneity , Phase cycling