• Title of article

    Improvement of hydrogen sorption measurement using a Sieverts apparatus with consideration of thermal transpiration

  • Author/Authors

    Cho، نويسنده , , Won Chul and Han، نويسنده , , Sang Sup and Kang، نويسنده , , Kyoung Soo and Bae، نويسنده , , Ki Kwang and Kim، نويسنده , , Change Hee and Jeong، نويسنده , , Seong-Uk and Park، نويسنده , , Chu Sik، نويسنده ,

  • Issue Information
    روزنامه با شماره پیاپی سال 2013
  • Pages
    8
  • From page
    6241
  • To page
    6248
  • Abstract
    Improvement of hydrogen sorption measurement using a Sieverts apparatus was performed by considering volumetric change due to the valve movement and the thermal transpiration. The newly suggested measurement method was evaluated by comparing the results from a blank test and a sample test with previous ones. The valve movement was found to be main measurement errors when the measurement temperature was near room temperature. The thermal transpiration became the major sources of errors as the difference between measurement temperature and room temperature increased. The modified equation for the PCI curve in this study gave the best performance compared with the previous ones at all measurement temperatures. These findings suggested the volumetric change due to thermal transpiration as well as the valve movement should be considered for more accurate measurements of hydrogen sorption.
  • Keywords
    Sievets apparatus , Thermal transpiration , Hydrogen measurement , Hydrogen , PCI curve
  • Journal title
    International Journal of Hydrogen Energy
  • Serial Year
    2013
  • Journal title
    International Journal of Hydrogen Energy
  • Record number