• Title of article

    Comparison of conforming and nonconforming retrieved glenoid components

  • Author/Authors

    Nho، نويسنده , , Shane J. and Ala، نويسنده , , Owen L. and Dodson، نويسنده , , Christopher C. and Figgie، نويسنده , , Mark P. and Wright، نويسنده , , Timothy M. and Craig، نويسنده , , Edward V. and Warren، نويسنده , , Russell F.، نويسنده ,

  • Issue Information
    روزنامه با شماره پیاپی سال 2008
  • Pages
    7
  • From page
    914
  • To page
    920
  • Abstract
    The purpose of this study was to compare differences in wear performance of conforming and nonconforming glenoid designs, using clinical, radiographic, and retrieved polyethylene glenoid component outcome Sixty-three glenoids met the study criteria, and each glenoid was assigned to the conforming group (if the radii of curvature of the humeral and glenoid components were identical) or the nonconforming group (if a mismatch existed between the radii of curvature). The mean length of implantation was 5.6 ± 5.5 years in the conforming group and 3.1 ± 3.1 years for the nonconforming group (P < .05). The loosening score was 3.2 ± 2.0 in the conforming group and 2.4 ± 1.2 in the nonconforming one (P < .05). The nonconforming group had a significantly greater burnishing score (P < .01), while the conforming group had greater abrasion and delamination scores (P < .05). Articular conformity contributes to differences observed from retrieved polyethylene glenoid components, which are consistent with differences in performance that may influence loosening.
  • Journal title
    Journal of Shoulder and Elbow Surgery
  • Serial Year
    2008
  • Journal title
    Journal of Shoulder and Elbow Surgery
  • Record number

    1867965