• Title of article

    Evaluation of the role of glenosphere design and humeral component retroversion in avoiding scapular notching during reverse shoulder arthroplasty

  • Author/Authors

    Berhouet، نويسنده , , Julien and Garaud، نويسنده , , Pascal and Favard، نويسنده , , Luc، نويسنده ,

  • Issue Information
    روزنامه با شماره پیاپی سال 2014
  • Pages
    8
  • From page
    151
  • To page
    158
  • Abstract
    Background ar notching is a common observation during radiological follow-up of reverse shoulder arthroplasty. The purpose of this study was to evaluate the effect of glenosphere design and humeral component retroversion on movement amplitude in the scapular plane and inferior scapular impingement. als and methods qualis Reversed Shoulder Prosthesis (Tornier) was implanted into 40 cadaver shoulders. On the glenoid side, 8 different combinations were tested: glenosphere: centered (standard), eccentric, with an inferior tilt, or with the center of rotation (COR) lateralized by 5 or 7 mm; and centered glenosphere: used alone or with the COR lateralized by 7 or 10 mm. meral component was positioned in 0°, 10°, 20°, 30°, and 40° of retroversion. Maximum adduction and abduction were measured when inferior impingement and superior impingement, respectively, were detected. s erage increase in abduction amplitude was 10° and inferior impingement occurred 18° later with a 42-mm glenosphere, especially when it was lateralized by 10 mm, relative to a 36-mm centered glenosphere (P < .05). These 2 combinations provided a 28° increase in the movement amplitude in the scapular plane. Positioning of the humeral component in 10° or 20° of retroversion or in anatomical retroversion was most effective at avoiding inferior impingement but had less effect on abduction range of motion (except with the 42-mm glenosphere). sion udy confirmed published results with various glenosphere designs but was unique in describing the effect of humeral retroversion on scapular impingement. Inferior scapular notching can be most effectively prevented by using large-diameter glenospheres with lateralized COR and by making sure to replicate the patientʹs native humeral retroversion.
  • Keywords
    reverse shoulder arthroplasty , glenoid modularity , humeral version , scapular notching
  • Journal title
    Journal of Shoulder and Elbow Surgery
  • Serial Year
    2014
  • Journal title
    Journal of Shoulder and Elbow Surgery
  • Record number