    Quantification of organic acids using voltammetric tongues

  • Author/Authors

    Escobar، نويسنده , , Juan David and Alcaniz، نويسنده , , Miguel and Masot، نويسنده , , Rafael and Fuentes، نويسنده , , Ana and Bataller، نويسنده , , Roman and Soto، نويسنده , , Juan and Barat، نويسنده , , Jose Manuel، نويسنده ,

  • Issue Information
    روزنامه با شماره پیاپی سال 2013
  • Pages
    7
  • From page
    814
  • To page
    820
  • Abstract
    Recently, electronic tongues (ET) have appeared as an excellent alternative to traditional techniques for the evaluation of food quality and processes. ET systems are based on arrays of low selectivity sensors that are simultaneously sensitive to several components in a measured sample (cross-sensitivity). The aim of this study was to determine the ability of an ET based on pulse voltammetry to quantify organic acids (ascorbic, citric and malic acids) in simple (SS) and binary solutions (BS) using different electrodes. The most significant electrodes for ascorbic acid prediction were Ni and Ag for SS, and Ag and Ir for BS where positive pulses were more suitable than negative ones. The prediction of citric and malic acids in SS and BS were suitable using Ir, Rh, Pt, Ag and Cu electrodes, using both positive and negative pulses.
  • Keywords
    PLS model , Binary solutions , organic acids , Pulse voltammetry , Cyclic voltammetry , Malic acid , Citric acid , ascorbic acid