• Title of article

    Coulometric Karl Fischer titration with a diaphragm-free cell: Cell design and applications

  • Author/Authors

    Lanz، نويسنده , , M. and De Caro، نويسنده , , C.A. and Rüegg، نويسنده , , K. and De Agostini، نويسنده , , A.، نويسنده ,

  • Issue Information
    روزنامه با شماره پیاپی سال 2006
  • Pages
    5
  • From page
    431
  • To page
    435
  • Abstract
    Coulometric Karl Fischer titration is a useful technique for wide-range water determination, from the percentage level down to trace amounts at the ppm level. Coulometric Karl Fischer titration cells having a current generator with a ceramic diaphragm, separating the anodic and the cathodic compartment, as well as diaphragm-free cells, are commercially available. From an applicative point of view, a diaphragm-free cell for Karl Fischer coulometry is favourable because of its ease-of-use. e found that improving the design of the diaphragm-free cell may contribute towards its application down to low water amounts. In this contribution, we will illustrate our recent work on optimising a diaphragm-free Karl Fischer cell for application over a wide water determination range, from the percentage level down to trace amounts at the ppm level.
  • Keywords
    Karl Fischer titration , Water determination , Traces , Diaphragm-free cell , Coulometry
  • Journal title
    Food Chemistry
  • Serial Year
    2006
  • Journal title
    Food Chemistry
  • Record number