• Title of article

    Evaluation of triple stage mass spectrometry as a robust and accurate diagnostic tool for determination of free cordycepin in designer egg

  • Author/Authors

    Chen، نويسنده , , Yi Hsin and Lim، نويسنده , , Chee Wei and Chan، نويسنده , , Sheot Harn، نويسنده ,

  • Issue Information
    روزنامه با شماره پیاپی سال 2014
  • Pages
    7
  • From page
    213
  • To page
    219
  • Abstract
    Direct determination of free cordycepin in designer egg using a highly selective mass spectrometric (MS) technique aided by a rapid and efficient dilute-and-shoot workflow would enhance their application as diagnostic tools in food fraud control. Here, triple stage mass spectrometry (MS3) demonstrated excellent analyte selectivity capability even when incomplete chromatographic separation was performed. Method validation was performed at six concentration levels of 100, 200, 400, 800, 1200 and 1600 ng g−1. Spiking experiments were examined at three concentration levels of 200, 400, and 1200 ng g−1 in individual egg white and egg yolk, measured over 2 days. MS3 enabled ion chromatograms with zero-background interference to be made in egg extracts. MS3 eliminated severe over recovery (p < 0.05) observed in all fortified samples, a challenge that MRM-transition could not address in a single step. Matrix-matched calibrants were needed to compensate for over recovery observed under MRM-transition mode.
  • Keywords
    Free cordycepin , Selectivity , Diagnostic , MS/MS/MS (MS3) , quantitation
  • Journal title
    Food Chemistry
  • Serial Year
    2014
  • Journal title
    Food Chemistry
  • Record number

    1976469