• Title of article

    UCN detection with 6Li-doped glass scintillators

  • Author/Authors

    Ban، نويسنده , , G. and Bodek، نويسنده , , K. and Lefort، نويسنده , , T. and Naviliat-Cuncic، نويسنده , , O. and Pierre، نويسنده , , E. and Plonka-Spehr، نويسنده , , C. and Rogel-Dorantes، نويسنده , , G.، نويسنده ,

  • Pages
    4
  • From page
    280
  • To page
    283
  • Abstract
    We report the results of test measurements aimed at determining the performance of 6Li-doped glass scintillators for ultra-cold neutron detection. Investigations have mainly focused on the reduction of the gamma-ray sensitivity of the scintillators. The probability of gamma interaction has been considerably lowered using very thin glasses. For signals corresponding to full-energy deposition, a background count rate of 8×10−3 s−1 was obtained for a shielded 0.5 cm3 GS10 scintillator located near the PF2 turbine at ILL. The neutron–gamma separation has further been improved using a stack with an 6Li-depleted scintillator and an 6Li-enriched one. Neutron captures leading to partial energy deposition (so-called “edge events”) have strongly been reduced resulting in a clear separation between the neutron and the gamma contributions.
  • Keywords
    6Li-doped glass scintillator , Efficiency of detection , Ultracold neutron
  • Journal title
    Astroparticle Physics
  • Record number

    1991317