• Title of article

    Energy dependent response of the Fricke gel dosimeter prepared with 270 Bloom gelatine for photons in the energy range 13.93 keV–6 MeV

  • Author/Authors

    Cavinato، نويسنده , , C.C. and Campos، نويسنده , , L.L.، نويسنده ,

  • Pages
    5
  • From page
    198
  • To page
    202
  • Abstract
    The spectrophotometric energy dependent response to photons with effective energies between 13.93 keV and 6 MeV of the Fricke xylenol gel (FXG) dosimeter developed at IPEN, prepared using 270 Bloom gelatine, was evaluated in order to verify the possible dosimeter application in other medicine areas in addition to radiosurgery, for example, breast radiotherapy and blood bags radiosterilization. Other dosimetric characteristics were also evaluated. The obtained results indicate that the FXG dosimeter can contribute to dosimetry in different medical application areas including magnetic resonance imaging (MRI) evaluation technique that permits three-dimensional (3D) dose distribution evaluation.
  • Keywords
    Fricke gel dosimeter , Spectrophotometry technique , Energy dependent response
  • Journal title
    Astroparticle Physics
  • Record number

    1991530