• Title of article

    Microtomography using a tube source of X-rays, a low-energy-resolution HPGe detector system and an array of detectors

  • Author/Authors

    Cesareo، نويسنده , , R and Rao، نويسنده , , D.V and Appoloni، نويسنده , , C.R and Brunetti، نويسنده , , A، نويسنده ,

  • Issue Information
    روزنامه با شماره پیاپی سال 1995
  • Pages
    6
  • From page
    573
  • To page
    578
  • Abstract
    A new microtomographic system has been assembled with a geometrical resolution of about 25–50 mm. This system is employed directly using the bremsstrahlung radiation and filtering the primary radiation with absorbers composed by an element. A reconstruction programme, written in PASCAL is able to give the reconstruction matrix of linear attenuation coefficients, simulates the matrix and the related tomography. Microtomography was studied thoroughly with different samples like acrylic cylinders containing internal W-wires of about 10 μm diameter, acrylic cylinders with holes of known diameters, small pencils with internal graphite, acrylic cylinders containing grains of soil and so on.
  • Journal title
    Nuclear Instruments and Methods in Physics Research Section A
  • Serial Year
    1995
  • Journal title
    Nuclear Instruments and Methods in Physics Research Section A
  • Record number

    1993693