• Title of article

    Elastic scattering induced peak tails in time-of-flight mass spectra

  • Author/Authors

    Ioanoviciu، نويسنده , , D.، نويسنده ,

  • Issue Information
    روزنامه با شماره پیاپی سال 1995
  • Pages
    4
  • From page
    454
  • To page
    457
  • Abstract
    Elastic scattering of an ion packet during its flight from source to detector generates peak tails on the high mass side in linear field-free space time-of-flight mass spectrometers, symmetrical, on both sides in reflectrons. The shape of these tails was calculated as a function of the elastic scattering cross section, the residual gas pressure and the detector size for linear drift space and symmetric reflectron time-of-flight mass spectrometers. rical example shows an abundance sensitivity several orders of magnitude better for a linear time-of-flight mass spectrometer than for a magnetic static deflector.
  • Journal title
    Nuclear Instruments and Methods in Physics Research Section A
  • Serial Year
    1995
  • Journal title
    Nuclear Instruments and Methods in Physics Research Section A
  • Record number

    1995693