Title of article
Setup optimization toward accurate ageing studies of gas filled detectors
Author/Authors
Y.P. and Abuhoza، نويسنده , , A. and Schmidt، نويسنده , , H.R. and Biswas، نويسنده , , S. and Frankenfeld، نويسنده , , U. and Hehner، نويسنده , , J. and Schmidt، نويسنده , , C.J.، نويسنده ,
Pages
3
From page
400
To page
402
Abstract
An infrastructure has been set up at the GSI detector laboratory to study the influence of construction materials on the ageing properties of gas filled detectors, such as multi-wire proportional chamber (MWPC), gas electron multiplier (GEM). Optimization of an ageing setup was performed by observing the variation of the normalized gain obtained using two identical MWPCs. An accuracy in the relative gain measurement below 1% has been achieved by monitoring environmental conditions and by systematic improvements of the measuring equipment. Ageing test of fiberglass G11 has been performed.
Keywords
MWPC , Gas detector , CBM , Mass Flow Controller (MFC) , Micro 3000 , RTV-3145 , Outgassing , ageing , Vetronit G11