• Title of article

    Experimental study of track density distribution on LR115 detector and deposition fraction of 218Po in diffusion chamber

  • Author/Authors

    Koo، نويسنده , , V.S.Y. and Yip، نويسنده , , C.W.Y. and Ho، نويسنده , , J.P.Y. and Nikezic، نويسنده , , D. and Yu، نويسنده , , K.N.، نويسنده ,

  • Pages
    4
  • From page
    470
  • To page
    473
  • Abstract
    The radial distribution of track density on a solid-state nuclear track detector inside a diffusion chamber is a function of the fraction f of 218Po decay before deposition. In the present work, procedures are proposed to determine f experimentally by determining the track density distribution on an LR115 detector in a diffusion chamber. First, a relatively tall diffusion chamber, with a height of 8 cm, was chosen. After exposure, the LR115 detector was etched. A transparent template with concentric circles was devised to study the radial distribution of sensitivity using an optical microscope. The distributions according to different values of f were also calculated using Monte Carlo simulations. By minimizing the deviations between these Monte Carlo curves and the experimental data, f was found to be 0.4.
  • Keywords
    natural radioactivity , RADON , Detector sensitivity , LR115 detector
  • Journal title
    Astroparticle Physics
  • Record number

    2020564