Title of article
Resolving pile up with a real-time processor for delay line detectors
Author/Authors
Descamps، نويسنده , , F.، نويسنده ,
Pages
8
From page
440
To page
447
Abstract
We report on the upgrade of a position-sensitive neutron detector for the D17 neutron reflectometer at the ILL Grenoble. The maximum count rate of this instrument has been increased to 420 kHz without sacrifice of spatial resolution. Using a multi-hit Time-to-Digital Converter and a RISC VME processor, we have implemented a real-time position reconstruction algorithm which improves the detector efficiency by recovering piled-up neutrons. The analysis relies on software cathode time checksums, thus avoiding position-dependent inefficiencies and reconstruction errors.
Keywords
Data acquisition electronics , neutron reflectometry , Pile up , Delay line PSD , Real-time processing
Journal title
Astroparticle Physics
Record number
2020980
Link To Document