Optimizing the design of a moderator-based neutron detector for a flat response curve in the 2–14 MeV energy range
Author/Authors
Hu، نويسنده , , Q.Y. and Li، نويسنده , , B.J. and Zhang، نويسنده , , D. and Guo، نويسنده , , H.S. and Yang، نويسنده , , G.Z. and Si، نويسنده , , F.N. and Liu، نويسنده , , J.، نويسنده ,
Pages
4
From page
213
To page
216
Abstract
The efficiency of a neutron detector with boron trifluoride proportional tubes embedded in a polyethylene moderator was simulated with a Monte Carlo program. A moderator structure where the detector had uniform sensitivity for neutrons from 2 to 14 MeV was determined by simulation. A counter was built based on the simulation results. The counterʹs efficiencies were calibrated with an Am–Be source and an accelerator that served as a D–D and D–T neutron source. Experimental neutron efficiencies of these sources are approximately uniform. The simulated model was validated by the consistent results between the calculated and experimental data.
Keywords
Monte Carlo simulation , Moderator , Energy response