• Title of article

    Time of flight (TOF) spectrometer for accurate measurement of mass and angular distribution of fission fragments in heavy ion induced fission reactions

  • Author/Authors

    Ghosh، نويسنده , , T.K. and Pal، نويسنده , , S. and Sinha، نويسنده , , T. and Chattopadhyay، نويسنده , , S. and Golda، نويسنده , , K.S. and Bhattacharya، نويسنده , , P.، نويسنده ,

  • Pages
    10
  • From page
    285
  • To page
    294
  • Abstract
    The performance of a dual time of flight (TOF) spectrometer using two 24 cm×10 cm area X – Y position sensitive Multi-Wire Proportional Counters (MWPC) is reported. Good separation of compound nuclear fission channel from elastic, quasi-elastic and non-compound fission channels have been achieved. Time calibration of the spectrometer using a technique, dependent on the detector independence of the mass or kinetic energy distributions gave a three times better mass resolutions compared to the conventional calibration technique using elastically scattered projectiles in a heavy ion induced fission reaction. Detail experimental methods in determining fission fragment mass and angular distribution have been presented and discussed.
  • Keywords
    Light heavy-ion reactions , mass , Angular distributions , MWPC
  • Journal title
    Astroparticle Physics
  • Record number

    2025772