• Title of article

    Fusion neutron detector calibration using a table-top laser generated plasma neutron source

  • Author/Authors

    Hartke، نويسنده , , R. and Symes، نويسنده , , D.R. and Buersgens، نويسنده , , F. and Ruggles، نويسنده , , L.E. and Porter، نويسنده , , J.L. and Ditmire، نويسنده , , T.، نويسنده ,

  • Pages
    6
  • From page
    464
  • To page
    469
  • Abstract
    Using a high intensity, femtosecond laser driven neutron source, a high-sensitivity neutron detector was calibrated. This detector is designed for observing fusion neutrons at the Z accelerator in Sandia National Laboratories. Nuclear fusion from laser driven deuterium cluster explosions was used to generate a clean source of nearly monoenergetic 2.45 MeV neutrons at a well-defined time. This source can run at 10 Hz and was used to build up a clean pulse-height spectrum on scintillating neutron detectors giving a very accurate calibration for neutron yields at 2.45 MeV.
  • Keywords
    Laser , Cluster , Fusion , Calibration , detector , Neutron
  • Journal title
    Astroparticle Physics
  • Record number

    2025820