• Title of article

    Modeling of Planck-high frequency instrument bolometers using non-linear effects in the thermometers

  • Author/Authors

    Piat، نويسنده , , Michel and Torre، نويسنده , , Jean-Pierre and Bréelle، نويسنده , , Eric and Coulais، نويسنده , , Alain and Woodcraft، نويسنده , , Adam K. Holmes، نويسنده , , Warren and Sudiwala، نويسنده , , Rashmi، نويسنده ,

  • Pages
    3
  • From page
    588
  • To page
    590
  • Abstract
    The Planck satellite, which is planned to be launched in 2007, is dedicated to surveying the Cosmic Microwave Background (CMB) to a high precision. Aboard this mission, the High-Frequency Instrument (HFI) will use 52 NTD Ge spiderweb bolometers made by Caltech-JPL and cooled to 100 mK by a dilution cooler. In this paper, we present a model of these detectors that includes non-linear effects seen in NTD Ge thermometers: electron–phonon decoupling and electrical field effect. We show that this model leads to consider only electrical field effect. Furthermore, the optical characterization of the HFI bolometers clearly shows a non-ideal behavior that is explained by non-linear effects in the thermometer. We finally show that these effects have to be taken into account for optimized CMB observations and to fully understand the physics of semi-conducting bolometers.
  • Keywords
    Spider web bolometer , Planck HFI , Electrical field effect , Electron–phonon decoupling effect
  • Journal title
    Astroparticle Physics
  • Record number

    2028316