• Title of article

    Characterization and quality control of avalanche photodiode arrays for the Clear-PEM detector modules

  • Author/Authors

    Abreu، نويسنده , , Suezilde da Conceiçمo do Amaral، نويسنده , , Pedro and Carriço، نويسنده , , Bruno and Ferreira، نويسنده , , Miguel and Luyten، نويسنده , , Joan and Moura، نويسنده , , Rui and Ortigمo، نويسنده , , Catarina and Rato، نويسنده , , Pedro and Varela، نويسنده , , Joمo، نويسنده ,

  • Pages
    4
  • From page
    19
  • To page
    22
  • Abstract
    Clear-PEM is a Positron Emission Mammography (PEM) prototype being developed in the framework of the Crystal Clear Collaboration at CERN. This device is a dedicated PET camera for mammography, based on LYSO:Ce scintillator crystals, Avalanche PhotoDiodes (APD) and a fast, low-noise electronics readout system, designed to examine both the breast and the axillary lymph node areas, and aiming at the detection of tumors down to 2 mm in diameter. ototype has two planar detector heads, each composed of 96 detector modules. The Clear-PEM detector module is composed of a matrix of 32 identical 2×2×20 mm3 LYSO:Ce crystals read at both ends by Hamamatsu S8550 APD arrays (4×8) for Depth-of-Interaction (DoI) capability. D arrays were characterized by the measurement of gain and dark current as a function of bias voltage, under controlled temperature conditions. Two independent setups were used. The full set of 398 APD arrays followed a well-defined quality control (QC) protocol, aiming at the rejection of arrays not complying within defined specifications. total of 398 arrays, only 2 (0.5%) were rejected, reassuring the trust in these detectors for prototype assembly and future developments.
  • Keywords
    quality control , Avalanche photodiode , Dark current , gain , Positron emission mammography
  • Journal title
    Astroparticle Physics
  • Record number

    2028727